Αρχειοθήκη ιστολογίου

Αναζήτηση αυτού του ιστολογίου

Τετάρτη 26 Δεκεμβρίου 2018

Image-guided study of inter-fraction and intra-fraction set-up variability and margins in reverse semi-decubitus breast radiotherapy

Abstract

Background

This study aimed to evaluate the inter-fraction set-up error and intra-fraction motion during reverse semi-decubitus (RSD) breast radiotherapy, and to determine a planning target volume (PTV) margin.

Material and methods

Pre- and post-treatment cone-beam computed tomography (CBCT) scans were prospectively acquired at fractions 1, 4, 7, 8, 11, and 14 for 30 patients who underwent RSD breast radiotherapy. Online correction for initial set-up error greater than 5 mm or 2° was performed and post-correction CBCT was acquired. An off-line analysis was performed to quantify initial and residual inter-fraction set-up errors and intra-fraction motion in three-dimensions. Patient inter-fraction errors were analysed for time trends during the course of radiotherapy. PTV margins were calculated from the systematic and random errors.

Results

The initial inter-fraction population systematic errors were 1.8–3.3 mm (translation) and 0.5° (rotation); random errors were 1.8–2.1 mm (translation) and 0.3–0.5° (rotation). After online correction, the residual inter-fraction population systematic errors were 1.2–1.8 mm (translation) and 0.3–0.4° (rotation); random errors were 1.4–1.6 mm (translation) and 0.3–0.4° (rotation). Intra-fraction population systematic and random errors were ≤ 1.3 mm (translation) and ≤ 0.2° (rotation). The magnitude of inter-fraction set-up errors in the anterior-posterior direction, roll, and yaw were significantly correlated with higher body weight and body mass index (BMI). The inter-fraction set-up error did not change significantly as a function of time during the course of radiotherapy. The magnitude of intra-fraction motion was not correlated with patient characteristics and treatment time. The total PTV margins accounting for pre-correction and intra-fraction errors were 6.5–10.2 mm; those accounting for post-correction and intra-fraction errors were 4.7–6.3 mm.

Conclusions

CBCT is an effective modality to evaluate and improve the inter-fraction set-up reproducibility in RSD breast radiotherapy, particularly for patients with higher BMI. Intra-fraction motion was minimal during RSD breast radiotherapy.



http://bit.ly/2Sk4ykI

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.