Abstract
Background and Purpose
The Barrel device is an electrolytically detachable laser cut, closed-cell microstent that is used for neck reconstruction in wide-necked bifurcation and branching aneurysms to support coiling. The key feature is a barrel section that herniates over the aneurysmal ostium. The objective was to evaluate the safety, feasibility and the immediate and mid-term occlusion results of this new device.
Material and Methods
The databases of two tertiary care centers were screened for all Barrel-based stent-assisted intracranial coil embolization of wide-necked aneurysms between June 2015 and September 2016. Case files and imaging data were retrospectively analyzed for angiographic and clinical outcome parameters, including immediate and mid-term modified Raymond-Roy aneurysm occlusion classification (RROC) rates and procedural complications.
Results
A total of 21 patients comprising 21 intracranial aneurysms (20 unruptured, 1 ruptured) were treated with the Barrel device and additional coiling of the aneurysm sac. All aneurysms were wide-necked, saccular bifurcation aneurysms defined by a dome/neck ratio ≤2. Immediate complete occlusion (RROC1) was observed in 19/21 (90%). An intra-interventional in-stent thrombus formation in two cases (10%) was medically resolved without neurological sequelae. A single case of symptomatic in-stent stenosis (5%) was cleared by balloon angioplasty. Follow-up (FU) was available in 20/21 cases (95%) after a median of 282 days (range: 17–591 days). At follow-up 19/20 aneurysms (95%) were completely occluded (RROC1).
Conclusion
The Barrel device showed a satisfactory safety profile and a promising rate of immediate and mid-term complete aneurysm occlusion for stent-assisted coil embolization of wide-necked intracranial bifurcation aneurysms, warranting further investigation of the device.
http://ift.tt/2mc8Mfy
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου
Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.