High Mass Ion Detection with Charge Detector Coupled to Rectilinear Ion Trap Mass Spectrometer

Abstract

Conventional linear ion trap mass analyzers (LIT-MS) provide high ion capacity and show their MSn ability; however, the detection of high mass ions is still challenging because LIT-MS with secondary electron detectors (SED) cannot detect high mass ions. To detect high mass ions, we coupled a charge detector (CD) to a rectilinear ion trap mass spectrometer (RIT-MS). Immunoglobulin G ions (m/z ~150,000) are measured successfully with controlled ion kinetic energy. In addition, when mass-to-charge (m/z) ratios of singly charged ions exceed 10 kTh, the detection efficiency of CD is found to be greater than that of SED. The CD can be coupled to LIT-MS to extend the detection mass range and provide the potential to perform MSn of high mass ions inside the ion trap.