Cyclic fatigue resistances of several nickel-titanium glide path rotary and reciprocating instruments at body temperature

Abstract

Aim

To compare the cyclic fatigue resistance of the One G, ProGlider, HyFlex EDM and R-Pilot glide path NiTi files at body temperature.

Methodology

Twenty One G (size 14, .03 taper), 20 ProGlider (size 16, .02 taper), 20 HyFlex EDM (size 10, .05 taper) and 20 R-Pilot (size 12.5, .04 taper) instruments were operated in rotation at 300rpm (One G, ProGlider and HyFlex) or in reciprocation (R-Pilot) at 35°C in artificial canals that were manufactured by reproducing the size and taper of the instrument until fracture occurred. The time to fracture was recorded in seconds using a digital chronometer and the length of the fractured fragments was registered. Mean data were analyzed statistically using the Kruskal-Wallis test and post-hoc Tukey tests via SPSS 21.0 software. The statistical significance level was set at 5%.

Results

The cyclic fatigue resistance of the R-Pilot files was significantly greater than the other instruments and the One G was significantly lower (p < 0.05). There was no difference between the HyFlex EDM and the ProGlider (p > 0.05). No significant difference (p > 0.05) was evident in the mean length of the fractured fragments of the various instruments.

Conclusions

The cyclic fatigue resistance of the R-Pilot reciprocating glide path file was significantly greater than that of the rotary HyFlex EDM, ProGlider and One G glide path files.

This article is protected by copyright. All rights reserved.



http://ift.tt/2BHp31L

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.