Αρχειοθήκη ιστολογίου

Αναζήτηση αυτού του ιστολογίου

Κυριακή 28 Φεβρουαρίου 2021

Robotic Posterior Capsule Polishing by Optical Coherence Tomography Image Guidance

xlomafota13 shared this article with you from Inoreader

Abstract

Background

In cataract surgery, polishing of the posterior capsule (PC) can lead to improved surgical outcomes but is currently avoided due to its high‐risk nature. This work developed a robotic system capable of performing PC polishing on ex vivo pig eyes using optical coherence tomography (OCT) guidance.

Methods

The lenses of five ex vivo pig eyes were extracted and a thin layer of glue deposited onto the PC. Transpupillary OCT scans of the anterior segment were used to generate a PC‐polishing trajectory. During polishing, OCT B‐scans tracked the tool tip and were displayed to the operator.

Results

Complete removal of the glue was accomplished in all five trials with no PC rupture reported.

Conclusions

The feasibility of using a robotic system guided by OCT to perform PC polishing on a biological model was demonstrated. Contributions include modeling of the PC anatomy, intraoperative OCT visualization, and automated tool‐tip motion with scheduled aspiration pressures.

This article is protected by copyright. All rights reserved.

View on the web

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.