Αρχειοθήκη ιστολογίου

Αναζήτηση αυτού του ιστολογίου

Τετάρτη 17 Αυγούστου 2016

Evaluation of Abdominal Computed Tomography Image Quality Using a New Version of Vendor-Specific Model-Based Iterative Reconstruction.

Purpose: To qualitatively and quantitatively compare abdominal computed tomography (CT) images reconstructed with a new version of model-based iterative reconstruction (Veo 3.0; GE Healthcare) to those created with Veo 2.0. Materials and Methods: This retrospective study was approved by our institutional review board and was Health Insurance Portability and Accountability Act compliant. The raw data from 29 consecutive patients who had undergone CT abdomen scanning was used to reconstruct 4 sets of 3.75-mm axial images: Veo 2.0, Veo 3.0 standard, Veo 3.0 5% resolution preference (RP), and Veo 3.0 20% RP. A slice thickness optimization of 3.75 mm and texture feature was selected for Veo 3.0 reconstructions. The images were reviewed by 3 independent readers in a blinded, randomized fashion using a 5-point Likert scale and 5-point comparative scale. Multiple 2-dimensional circular regions of interest were defined for noise and contrast-to-noise ratio measurements. Line profiles were drawn across the 7 lp/cm bar pattern of the CatPhan 600 phantom for spatial resolution evaluation. Results: The Veo 3.0 standard image set was scored better than Veo 2.0 in terms of artifacts (mean difference, 0.43; 95% confidence interval [95% CI], 0.25-0.6; P

http://ift.tt/2bFQUrL

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.